Model Number 21-70005
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/10/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Failure of osseointegration.
 
Additional Manufacturer Narrative
Failure to osseointegrate is a well-known and inherent risk of dental implants and is documented in the literature across implant systems 1, 2 (see e.G., keller, 1995; bornstein, 2008,).Paltop dental solutions inc.Provides ifu with all implants.Failure to osseointegrate, and loss of osseointegration (implant mobility) is identified as an adverse reaction and identified in the list of warnings in all end-osseous dental implant labelling.The specific cause for a particular complaint is often not readily identified as there are various factors that contribute to the risk of implant failure.These include systemic medical condition (uncontrolled diabetes mellitus, aids, osteoporosis), poor bone quality and quantity of bone in which the implant is implanted, medications (corticosteroids, bisphosphonates), harmful habits to healing (smoking), secondary infection and/or surgical trauma, unexperienced clinician, improper surgical technique, or mobility of the implant leads to its removal or replacement, hence dental implant failure.In addition, a rate of implant failure that does not affect the products' risk assessment does not require investigation as it is a well-known failure mode which has been adequately studied in the past.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DYMIC
Common Device Name
DYNAMIC DENTAL IMPLANT Ø4.2MM L6MM
Manufacturer (Section D)
PALTOP ADVANCED DENTAL SOLUTIONS INC.
13645 alton pkwy ste a
irvine CA 92618
Manufacturer (Section G)
PALTOP ADVANCED DENTAL SOLUTIONS INC.
13645 alton pkwy ste a
irvine CA 92618
Manufacturer Contact
brian lanza
13645 alton pkwy ste a
irvine, CA 92618
7742100390
MDR Report Key25454977
Report Number3009161350-2026-01493
Device Sequence Number10450245
Product Code DZE
UDI-Device Identifier07290108695496
UDI-Public(01)07290108695496(17)280416(10)WO-025959
Combination Product (Y/N)N
Initial Reporter CountryRS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 06/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number21-70005
Device Catalogue Number21-70005
Device Lot NumberWO-025959
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/10/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2026
Initial Report FDA Received Date06/10/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/23/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-