During an endoscopic variceal ligation in the esophagus, the physician used a cook 6 shooter saeed multi-band ligator.It was reported that the physician installed the device normally, after aspiration the physician released the device and found that the released white band stuck under band and the endoscope cannot be retracted.After releasing all the rubber rings [bands] at once, the endoscope can be retracted.Other than the deployed bands which will pass naturally, a section of the device did not remain in the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation evaluation: the product said to be involved was returned in a white plastic bag with an open box and tray from the lot number provided in the report.The label matches the product returned.The loading catheter, irrigation adapter, and bands did not return with the device.Photos were provided and reviewed.The first photo was of the box label, and the lot number matches in the report.The second photo is an endoscopic image, the end of the barrel can be seen, and two bands are stuck on the trigger cord.And attempted to be deployed on the varices.The third photo shows all of the bands deployed in the patient, as they stated, to remove the endoscope from the patient.Our laboratory evaluation of the product said to be involved confirmed the report as it was described, the trigger cord was no longer attached to the barrel, and the trigger cord was wrapped around the handle in a post deployment state.A visual examination of the device was performed.The trigger cord was examined and all twelve deployment beads were present.The beads were examined using magnification and found to be correctly filled and had no evidence of excess flash.The trigger cord was intact and not broken.The length of the trigger cord was measured at 143.1 cm between the knots which is within tolerance.An evaluation of the handle wheel movement was performed and the wheel functioned as intended.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our laboratory evaluation of the returned device confirmed the report based on the condition of the returned device, all bands deployed.A definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Band deployment difficulty can occur if the endoscope accessory channel is compromised.In these cases, the endoscope accessory channel collapses, restricting the trigger cord and preventing proper band deployment.The instructions for use contain the following statement: "use of an endoscope in a sound state of repair is a prerequisite for a successful multi-band ligation procedure." band deployment difficulty can also occur if the trigger cord is not properly seated in the handle assembly.The instructions for use advise the user: "note: knot must be seated into hole or handle will not function properly." a precaution in the instructions for use state: "it is vital that the integrity of the working channel is intact as grooves or other obstructions in the working channel can potentially cause the string to catch, resulting in band deployment difficulty." it is possible that this could lead to band deployment difficulty and damage to the trigger cord as well.Prior to distribution, all 6 shooter saeed multi-band ligators are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|