| Model Number |
429536 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
Ventricular Fibrillation (2130)
|
| Date of Event |
05/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
During rf ablation with medtronics affera ablation system, the electrical output of the sphere-9 ablation catheter, interacted with this device.The energy caused the patient to go into ventricular fibrillation twice.The patient was successfully shocked out of this rhythm.Should additional information become available, this file will be updated.Uf/importer report #: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Biotronik submits this report to comply with fda regulations 21 cfr parts 4 and 803.Biotronik has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by biotronik, or its employees that the device, biotronik, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Biotronik will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn based on available information at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|