Model Number 9600TFX29A
Medical Device Problem Codes Calcified (1077); Perivalvular Leak (1457)
Health Effect - Clinical Codes Dyspnea (1816); Fatigue (1849)
Date of Event 05/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
As reported by an edwards lifesciences field clinical specialist, a 29mm sapien 3 valve, implanted in the aortic position, failed due to significant paravalvular leak (pvl) approximately 5 years and 11 months post implant.The gradient was reported as 35mmhg, but it is unknown if it was a peak or mean valve gradient.The perceived root cause of the pvl was thought to be bulky calcium at the annulus.The patient presented with fatigue and shortness of breath.A valve-in-valve procedure was performed, and a 26mm sapien 3 ultra resilia valve was successfully implanted within the sapien 3 valve.The patient is stable and in recovery.
 
Additional Manufacturer Narrative
Per the instructions for use (ifu), paravalvular leak (pvl) is a known potential adverse event associated with bioprosthetic heart valves and the transcatheter valve replacement (tvr) procedure.Paravalvular leak refers to blood flowing through a channel between the structure of the implanted valve/pre-stent and the landing zone, as a result of a lack of appropriate sealing of the valve to the target site.Some pvl is not uncommon post-deployment.Many cases are mild to moderate, and either resolve over time or do not cause symptoms.Others may be more clinically significant and require intervention.The patient screening manual and the procedure didactic identify several procedural and anatomical factors which could contribute to pvl, including device malposition, inaccurate measurement of the landing zone, uneven distribution of calcium on the landing zone, bulky or severe calcification, an elliptical annulus shape, and valve under-sizing.The mechanism behind worsening or late pvl is not well understood but may be related to cardiac remodeling.Edwards extensively trains physicians before they are qualified to use the sapien transcatheter heart valves (all models).The device training manuals instructs the operator on proper imaging screening requirements, including the use of good quality echocardiography and/or computed tomography (ct) to appropriately measure the landing zone, assess the content and distribution of calcium, and other patient anatomical factors.Multiple imaging modalities should be considered during valve size selection.Contraindications, important considerations when assessing the valve, and choosing the proper transcatheter heart valve are also discussed.The device training manuals also instruct the operator on proper positioning and deployment of the valve, including patients screening and procedural considerations.Device preparation, approach, deployment, imaging, procedure-specific training manuals, and proctored procedures are also included.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results indicate patient factors (bulky calcium) caused or contributed to this event.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAPIEN 3 TRANSCATHETER HEART VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key25460581
Report Number2015691-2026-15930
Device Sequence Number18754448
Product Code NPT
UDI-Device Identifier00690103194364
UDI-Public(01)00690103194364(17)220119
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P140031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/19/2022
Device Model Number9600TFX29A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2026
Supplement Date Received by Manufacturer06/11/2026
Initial Report FDA Received Date06/10/2026
Supplement Report FDA Received Date07/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/20/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age81 YR
Patient SexMale
-
-