| Model Number |
P7800A |
| Medical Device Problem Code |
Activation Problem (4042)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Baxter received a report that compella cap us scale pd ppm had cpr function not activated.This occurred during baxter servicing/testing.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The baxter technician found the cpr handle needed to be replaced.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in jul 21, 2025.It is unknown if the facility performed any other preventative maintenance on this bed.Per the baxter service manual the compella bed requires an effective maintenance program.We recommend that you perform annual preventative maintenance.Pay particular attention to safety features, which include but are not limited to: all caregiver and patient controls, along with their indicators, operate correctly; this includes controls for cpr.Replace or repair parts as necessary.The technician replaced the cpr handle to resolve the reported event.Based on this information, no further action is required.
|
| |
|