ALLERGAN (COSTA RICA) Inspira Smooth Silicone Gel Filled Breast Implant; prosthesis, breast, noninflatable, internal, silicone gel-filled Back to Search Results
Catalog Number SSLP-300
Medical Device Problem Codes Device Appears to Trigger Rejection (1524); Material Rupture (1546)
Health Effect - Clinical Codes Capsular Contracture (1761); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/09/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
 
Event or Problem Description
Healthcare professional reported "stable intracapsular implant leak".This record is for the right side.The device remains implanted.
 
Event or Problem Description
Healthcare professional reported "stable intracapsular implant leak".Later healthcare professional reported "hole, non-specific", "gel bleed" and capsular contracture baker grade unknown.This record is for the right side.The device was explanted.
 
Event or Problem Description
Later healthcare professional reported "hole, non-specific", "gel bleed" and capsular contracture baker grade unknown.The device was explanted.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: b5, d9, h3, h6.Device evaluation: per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: - rupture: observed an opening through microscopic inspection assessed as surgical damage.No further actions are required as it is not related to the manufacturing process.- capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (crease) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
 
Additional Manufacturer Narrative
Additional, changed, and/or corrected data: a4, a6, b5, d6b, h6.This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
Inspira Smooth Silicone Gel Filled Breast Implant
Common Device Name
prosthesis, breast, noninflatable, internal, silicone gel-filled
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
la aurora de heredia
CS 
Manufacturer (Section G)
Allergan (Costa Rica)
900 parkway global park
zona franca
la aurora de heredia
CS  
Manufacturer Contact
terry ingram
2525 dupont drive
irvine, CA 92612
+1(847)936-6324
MDR Report Key25461722
Report Number9617229-2026-11026
Device Sequence Number16054887
Product Code FTR
UDI-Device Identifier10888628032736
UDI-Public(01)10888628032736(17)221120
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/20/2022
Device Catalogue NumberSSLP-300
Device Lot Number3143028
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer09/03/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/14/2026
Supplement Date Received by Manufacturer07/02/2026
09/18/2026
Initial Report FDA Received Date06/10/2026
Supplement Report FDA Received Date07/27/2026
10/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/20/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient SexFemale
Patient Weight121 LB
Patient RaceWhite
-
-