| Model Number |
CF-H170I |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, it is possible that high-energy endo therapy accessories (lasers, radiofrequency, etc.) were used without maintaining sufficient distance from the tip resulting in the malfunction.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the colonovideoscope to the service center for evaluation related to a separate issue.During testing, the service repair technician identified the device had burnt c-cover.There was no patient involvement.
|
| |
|