Product complaint #(b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.
|
|
It was reported on (b)(6) 2026, that when final tightening 2 trialtis set screws completely stripped threads out in same screw tulip head, 3rd set screw/3rd attempt worked.This was a sai screw 8x65mm, no counter torque was used because the screw was buried in bone and counter torque could not fit over tulip head.5 minutes surgery delayed due to the reported event.Tried again with new set screw.Procedure was successfully completed.Fragments were generated and they removed easily without additional intervention.No patient consequences.
|