| Model Number |
6487800 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
On-site investigation was performed by distributor's field service engineer.The claimed issue was reproduced during troubleshooting.Replacement of the pressure transducer printed circuit board (pcb) solved the issue.The ventilator passed all functional and safety tests and was cleared for clinical use.Pressure transducer pcb measures the pressure, conveyed via the pressure tube connected to this block by its differential pressure transducer.With differential reference to the ambient pressure, the output signal is proportional to the measured pressure thus giving a linear measurement in the range -40 cmh2o to +160 cmh2o.The device's logs were not provided for further analysis.The defective part was not returned for investigation, despite request.The cause of the reported issue in this customer product complaint has been determined.The issue was related to pressure transducer pcb.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator failed internal leakage test during pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
|
| |
|
Search Alerts/Recalls
|