| Model Number |
PM700 |
| Medical Device Problem Codes |
Loss of Power (1475); Noise, Audible (3273)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3:product analysis :evaluation couldn't able to reproduce the reported malfunction of loss of power.It was also found that hose detached from the swivel knuckle due to missing clamp ring and 3 small balls of the standoff swivel was missing.¿this regulatory report is being submitted due to retrospective review through capa 672570.¿ medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the motor with the report of loss of power and noise.It was reported that there was no patient impact.Additional information was received stating that hose detached were found during evaluation.
|
| |
|
Search Alerts/Recalls
|