Model Number 40-9366
Medical Device Problem Code Complete Blockage (1094)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a nasogastric (ng) tube became clogged during use and could not be flushed, requiring removal of the tube.No injury or medical intervention was reported.
 
Additional Manufacturer Narrative
This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.A review of the device history record is not possible as no lot number was provided; therefore, the udi-pi is unavailable.The actual complaint product was returned for evaluation.There is no visible occlusion.Feeding was simulated using water.A 35ml enfit syringe filled with water was attached to the central port.Upon infusion, the water flowed without resistance or obstruction, exiting the bolus tip.No failures were detected.The device functioned as intended.The complaint has not been confirmed.All information reasonably known as of 11 jun 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as (b)(4).This information is submitted pursuant to 21 cfr 803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
tamara cardona
5405 windward parkway
alpharetta, GA 30004
4704485487
MDR Report Key25471903
Report Number9611594-2026-00431
Device Sequence Number18758360
Product Code KNT
UDI-Device Identifier00350770460420
UDI-Public(01)00350770460420
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K821906
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number40-9366
Device Catalogue Number40-9366
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2024
Initial Report FDA Received Date06/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-