• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: UNOMEDICAL DEVICES S.A DE C.V INSET II; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL DEVICES S.A DE C.V INSET II; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number MMT-943A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2025
Type of Reportable Event Death
Additional Manufacturer Narrative
Unomedical hereby submits this initial report as part of its complaint remediation activities conducted under a capa 2356421 and capa 2566479 in accordance with the fda action plan, which incorporates qp-ic-2026-0020 (ic retrospective complaint review) and qp-ic-2026-0024 (ic retrospective mdr review).This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this supplemental information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged.Additional information - this medical device report (mdr) is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions.H11: investigation summary; complaint investigation results: a complaint investigation was performed based on the event description and no malfunction based on complaint information.Additional information was requested to support the investigation; however, a lot number or any product specific information was not provided.No device, device components, or other visual or physical evidence was made available for evaluation.Consequently, visual inspection, retain-sample testing, or the assessment of potential product performance issues, component integrity, or manufacturing defects could not be performed.In response to the complaint and due to the absence of a specific lot number, a high-level review of manufacturing controls for the minimed mio product family was conducted to document the routine controls used to detect potential defects at the product family level.The review identified the following controls in use for the minimed mio product family, including: 100% leak and flow testing during final assembly 100% visual inspections during packaging, verifying needle condition, tyvek seal integrity, contamination-free condition, and correct assembly.Sampling verification under acceptable quality limit (aql) 0.65, including visual and functional tests such as: burst test peel test flow and leak tests static tension test visual inspections for cannula, tubing, lid, heat-shrink and assembly integrity.Corrective and preventive action (capa) determination: based on the investigation, no further action is required.The record will be closed and monitored through tracking and trending per work instruction (wi) (monthly trips and alerts).Complaint investigation - conclusion: based on the limited information available, the investigation was completed and no product malfunction was alleged or identified.The complaint could not be confirmed due to insufficient information, and the record is being closed based on the event description and the information provided in the complaint.Based on the available information, this event is deemed to be a reportable death.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 8021545 manufacturing site: 8021545.
 
Event or Problem Description
It was reported that patient died on (b)(6) 2025 and the cause of the death is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSET II
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
UNOMEDICAL DEVICES S.A DE C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX  88736
Manufacturer (Section G)
UNOMEDICAL DEVICES S.A DE C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX   88736
Manufacturer Contact
courtney smith
aaholmvej 1 - 3
osted
lejre 4320
DA   4320
6032470255
MDR Report Key25472118
Report Number3003442380-2026-31708
Device Sequence Number16058722
Product Code FPA
UDI-Device Identifier05705244020030
UDI-Public05705244020030
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-943A
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2026
Initial Report FDA Received Date06/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexUnknown
-
-