MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION CAREASSIST BED; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION CAREASSIST BED; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Model Number P1170C
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/15/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The baxter technician found all four casters needed to be replaced.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in may 1, 2026.It is unknown if the facility performed any other preventative maintenance on this bed.Per the baxter service manual the bed should be subject to an effective maintenance program.An annual service of the bed is advised in order to maintain its characteristics and performance.Brake casters should be checked for cuts, wear and quality of tread, etc.And replaced when necessary.Check the brakes to see whether the bed moves when the brake pedals are pressed and repair as necessary.The technician replaced all four casters to resolve the reported event.Based on this information, no further action is required.
 
Event or Problem Description
Baxter received a report that care assist ca305 had brakes not holding while in brake position.The process step during which this occurred was unknown.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAREASSIST BED
Common Device Name
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
HILL-ROM MEXICO
ave.del telefono
no. 200 col. huinala
apodaca, nuevo leon 66640
MX   66640
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242705355
MDR Report Key25474375
Report Number3006697241-2026-00218
Device Sequence Number16040511
Product Code FNL
UDI-Device Identifier00887761000278
UDI-Public(01)00887761000278
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberP1170C
Device Catalogue NumberP1170C0000008
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/15/2026
Initial Report FDA Received Date06/11/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-