Catalog Number 035.7008S
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Fibrosis (3167)
Date of Event 05/08/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The removal of a dental implant during surgery without the replacement of another dental implant is a known inherent risk of the procedure due to either lack of primary stability of the implant (patienor procedure related).It may also include the removal of an implant after osseointegration due to either the clinician's or patient¿s decision.The manufacturer¿s trend analysis confirms that usually procedural errors and/or patient's condition contribute to the event.
 
Event or Problem Description
The clinician reports that the day the implant was placed in ada 30, failure occurred upon insertion.The device was forwarded to the manufacturer.At the event the patient experienced: mobility.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TLC SP, Ø3.3MM NT, SLACTIVE® 8MM RXD
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ  04002
Manufacturer (Section G)
INSTITUT STRAUMANN AG
peter merian-weg 12
basel 04002
SZ   04002
Manufacturer Contact
jennifer jackson
60 minuteman road
andover, MA 01810
9787472509
MDR Report Key25475264
Report Number0009613348-2026-008884
Device Sequence Number10465950
Product Code DZE
UDI-Device Identifier07630031766045
UDI-Public07630031766045
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K230108
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number035.7008S
Device Lot NumberVJYX1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/11/2026
Event Location Other
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR