| Catalog Number |
0475100000 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/14/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the crossfire 2 console caught on fire mid-procedure after rf wand was activated.
|
| |
|
Event or Problem Description
|
|
It was reported that the crossfire 2 console caught on fire mid procedure after rf wand was activated.
|
| |
|
Additional Manufacturer Narrative
|
|
Product was not received in-house at stryker endoscopy.This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: console caught on fire mid procedure:
probable root cause:
console fuses fail to break circuit due to excessive mains current leading to fire,
isolation power supply failure,
use error: console is sterilized or disinfected,
use error: console cleaned with incompatible products.The reported failure mode will be monitored for future reoccurrence.
|
| |
|