• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE; PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 320-10-00
Medical Device Problem Codes Unstable (1667); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Joint Laxity (4526); Metal Related Pathology (4530)
Date of Event 05/20/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient, who had a left tsa, underwent a revision procedure.The patient presented with an unstable shoulder and required surgery to reduce the joint stability.During the case, it became clear of what the issue was as the poly was floating around in the joint space and was not locked into the humeral tray.There was lots of black tissue and metallosis surrounding the implants.The 38mm glenosphere and glenosphere locking screw were explanted along with the 38mm +2.5mm constrained liner, +0mm humeral tray and torque screw.They upsized and reimplanted a 42mm glenosphere and glenosphere locking screw, and a 42mm +2.5mm liner, +0mm humeral tray and torque screw.The glenoid baseplate and humeral stem were left in as they were well fixed.There were no device breakages or surgical delays during the procedure.The patient was last known to be in stable condition following the event.X-ray images were provided.The explanted devices are not available for return as the hospital won't release them.Device images were provided.No further information.This is 1 of 3 reports for this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE
Common Device Name
PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainseville FL 32653
Manufacturer Contact
julia goldberg
2320 nw 66 court
gainseville, FL 32653
3523771140
MDR Report Key25479814
Report Number1038671-2026-00765
Device Sequence Number14813776
Product Code KWT
UDI-Device Identifier10885862086419
UDI-Public10885862086419
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K063569
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number320-10-00
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2026
Supplement Date Received by Manufacturer06/24/2026
Initial Report FDA Received Date06/11/2026
Supplement Report FDA Received Date06/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/13/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age71 YR