ABBOTT MEDICAL MITRACLIP G5 DELIVERY SYSTEM; MITRAL VALVE REPAIR DEVICES Back to Search Results
Model Number CDS0801-XT
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The clip remains in the patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported on (b)(6) 2026 that the patient presented with grade 4 degenerative mitral regurgitation (mr) and prolapsed anterior leaflet for a mitraclip procedure.During the procedure, upon turning the m knob, the sleeve showcased unintended movement.The clip was removed from the anatomy.The same issue was observed in three more clips.However, the three clips were deployed successfully.The mr was reduced to grade 3-4 post procedure.There was no clinically significant delay or adverse patient effects reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G5 DELIVERY SYSTEM
Common Device Name
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL NORTHPOINT REG 3020950818
1820 bastian court
westfield IN 46074
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key25485356
Report Number2135147-2026-03962
Device Sequence Number18770019
Product Code NKM
UDI-Device Identifier05415067050762
UDI-Public(01)05415067050762(17)270518(10)51117A1061
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P100009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCDS0801-XT
Device Catalogue NumberCDS0801-XT
Device Lot Number51117A1061
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2026
Supplement Date Received by Manufacturer06/23/2026
Initial Report FDA Received Date06/12/2026
Supplement Report FDA Received Date06/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/18/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
STEERABLE GUIDE CATHETER.
Patient SexUnknown
-
-