MAUDE Adverse Event Report: KERI MEDICAL SA TOUCH - CMC1 PROSTHESIS; TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 8MM - M

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KERI MEDICAL SA TOUCH - CMC1 PROSTHESIS; TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 8MM - M Back to Search Results
Model Number NTO158
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Implant Pain (4561); Joint Impingement (4834)
Date of Event 05/19/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation related to this event is ongoing.At this time, the available information is insufficient to determine a definitive root cause or whether the device contributed to the reported event.A supplemental mdr will be submitted to fda upon completion of the investigation and when additional information becomes available.
 
Event or Problem Description
It was reported that a patient underwent a total joint arthroplasty of touch cmc1 prosthesis on (b)(6) 2024.Subsequently, the patient experienced pain and a grating sensation when moving the thumb, and underwent a revision surgery on (b)(6) 2026, due to breakage of the pe liner.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TOUCH - CMC1 PROSTHESIS
Common Device Name
TOUCH® - CMC1 PROSTHESIS OFFSET NECK 15° - 8MM - M
Manufacturer (Section D)
KERI MEDICAL SA
chemin du pre-fleuri 5
plan-les-ouates, 1228
SZ  1228
Manufacturer (Section G)
KERI MEDICAL FRANCE
34 rue antoine redier
archamps, 74160
FR   74160
Manufacturer Contact
camille miletto
chemin du pre-fleuri 5
plan-les-ouates, 1228
SZ   1228
MDR Report Key25486378
Report Number3024706750-2026-00049
Device Sequence Number9930006
Product Code SFA
UDI-Device Identifier07640181160051
UDI-Public07640181160051
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P240020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Remedial Action Replace
Type of Report Initial,Followup
Report Date (Section B) 07/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberNTO158
Device Catalogue NumberNTO158
Device Lot NumberX7821
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2026
Supplement Date Received by Manufacturer06/30/2026
Initial Report FDA Received Date06/12/2026
Supplement Report FDA Received Date07/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/17/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
-
-