• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. TRILOGY 200; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number CA1032800
Medical Device Problem Codes Circuit Failure (1089); Failure to Calibrate (2440)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/23/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The reported failure of nurse call issue is accommodated in the product risk management file as being linked to hazard with description loss of function/loss of primary function, and patient exposure to function / deterioration of function.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.The device mentioned in this complaint has exceeded its useful life per the applicable ifu.
 
Event or Problem Description
A trilogy 200 was returned to a third-party service center for service.There was no serious patient harm or injury that occurred or was reported.There was no allegation of device malfunction.The device was not in patient use.During the evaluation of the device at the third-party service center, the device failed a test step related to the nurse call function during testing.The device's interface board was replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 200
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key25486793
Report Number2518422-2026-021218
Device Sequence Number18570998
Product Code CBK
UDI-Device Identifier00606959022782
UDI-Public(01)00606959022782
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K093416
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 06/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberCA1032800
Device Catalogue NumberCA1032800
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/23/2026
Initial Report FDA Received Date06/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/11/2014
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-