• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY SPEEDPLATE - MICROQUAD; STAPLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TREACE MEDICAL CONCEPTS, INC. LAPIPLASTY SPEEDPLATE - MICROQUAD; STAPLE Back to Search Results
Model Number SK62
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that after an initial bunion surgery on (b)(6) 2025, hardware was removed on (b)(6) 2026 due to the hardware being prominent.There were no deficiencies or malfunctions alleged/found with any tmc device and no other patient outcomes were reported as a result of this event.
 
Additional Manufacturer Narrative
It was reported that after an initial bunion surgery on (b)(6) 2025, hardware was removed on (b)(6) 2026 due to the hardware being prominent.There were no deficiencies or malfunctions alleged/found with any tmc device and no other patient outcomes were reported as a result of this event.Additional information indicates the patient's bones were completely fused/healed.No devices were returned for evaluation.The device history records were reviewed and no issues were identified during the manufacture and release of the devices that could have contributed to what was reported.Although several factors could have contributed to what was reported, the most likely cause cannot be determined based on the available information and without return of the device.However, it is possible initial placement of the plate contributed to what the patient experienced.The company will supplement this mdr as necessary and appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAPIPLASTY SPEEDPLATE - MICROQUAD
Common Device Name
STAPLE
Manufacturer (Section D)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer (Section G)
TREACE MEDICAL CONCEPTS, INC.
100 palmetto park place
ponte vedra FL 32081
Manufacturer Contact
uriza shums
100 palmetto park place
ponte vedra, FL 32081
9043735940
MDR Report Key25488059
Report Number3011623994-2026-00146
Device Sequence Number16082313
Product Code JDR
UDI-Device Identifier00810111223645
UDI-Public(01)00810111223645(17)291205(10)300608823
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K251135
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSK62
Device Lot Number300608823
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2026
Initial Report FDA Received Date06/12/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/05/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-