(b)(4).Date sent: 6/12/2026.D4: batch #609d78.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the ec60a device was received for analysis with joint cover damaged and with a cecr60b reload loaded on the device.The reload was received partially fired 1/5.The device was returned with the jaws and closure trigger in the closed position.Also, the knife was noted partially advanced, the red manual knife reverse button was activated and the knife returned to the home position as expected.The device was tested for functionality in the articulated position with the returned reload by resetting and reloading it into the device.The device achieved its complete stroke firing sequence without any difficulties.The staple line and cut line were complete and the remaining staples met the staple form release criteria.The partially fired is consistent with an incomplete or interrupted cycle.The damaged on the joint cover, it is possible that the device was attended to be pulled out from a trocar in the articulated position, resulting in the edge of the trocar damaging the joint cover.As part of ethicon ¿s quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the device lot e25110057, and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 609d78, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|