| Catalog Number |
68HP-425 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "deflation".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device related to the reported events deflation was received on july 20, 2026, with lot number 3419784.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification (if applicable).Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ deflation: observed an opening through microscopic inspection assessed as unidentified (tear) opening.No further actions are required as it is not related to the manufacturing process.No additional observations performed on the device.No further actions are required.Additional, changed, and/or corrected data: d.9., h.2., h.3., h.6.
|
| |
|