| Model Number |
AA07 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3: product analysis: evaluation determined that the reported malfunction of overheating could not be able to confirmed.However it was also noted that the bearings were broken and laser markings were worn.B3: date of incident or estimated date of incident was not provided to the manufacturer.Event notified date used as date of incident until further information is obtained.G3: aware date used as date of analysis performed date as the event is reported based on evaluation findings.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during the procedure there was overheating observed in the attachment.At the time of report there was no patient impact.Additional information received upon evaluation stating that bearings were broken inside the tube.
|
| |
|
Search Alerts/Recalls
|