MAUDE Adverse Event Report: UNOMEDICAL DEVICES S.A DE C.V CONTACT DETACH; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

UNOMEDICAL DEVICES S.A DE C.V CONTACT DETACH; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number MMT-864
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/31/2020
Type of Reportable Event Death
Event or Problem Description
It was reported that patient was hospitalized and died on (b)(6) 2020 and the cause of death is heart attack.
 
Additional Manufacturer Narrative
Unomedical hereby submits this initial report as part of its complaint remediation activities conducted under a capa 2356421 and capa 2566479 in accordance with the fda action plan, which incorporates qp-ic-2026-0020 (ic retrospective complaint review) and qp-ic-2026-0024 (ic retrospective mdr review).This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this supplemental information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged additional information - this medical device report (mdr)is being submitted to include the below: h6: investigation results under type of investigation, investigation findings, investigation conclusions h11: investigation summary based on the available information, this event is deemed to be a reportable death.Request was performed for additional information including lot number; however, lot number was not provided.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 8021545.Manufacturing site: 3003442380.
 
Additional Manufacturer Narrative
Unomedical hereby submits this supplemental report as part of its complaint remediation activities conducted under a capa 2356421 and capa 2566479 in accordance with the fda action plan, which incorporates qp-ic-2026-0020 (ic retrospective complaint review) and qp-ic-2026-0024 (ic retrospective mdr review).This submission includes a retrospective reassessment of previously evaluated complaints.Unomedical is providing this supplemental information in accordance with the reporting requirements set forth in 21 cfr part 803.Any fields left blank indicate that the information is unknown, unavailable, or remains unchanged.Additional information - this medical device report (mdr)is being submitted to include the below: d2a: common device name h6: investigation results under type of investigation, investigation findings, investigation conclusions h11: investigation summary complaint investigation results a complaint investigation was performed based on the event description and the assigned malfunction code no malfunction based on complaint information.Additional information was requested to support the investigation; however, a lot number or any product specific information was not provided.No device, device components, or other visual or physical evidence was made available for evaluation.Consequently, visual inspection, retain-sample testing, or the assessment of potential product performance issues, component integrity, or manufacturing defects could not be performed.In response to the complaint and due to the absence of a specific lot number, a high-level review of manufacturing controls for the steel (sure-t paradigm (p-cap)) product family was conducted to document the routine controls used to detect potential defects at the product family level.The review identified the following controls in use for the steel (sure-t paradigm (p-cap) product family, including: 100% visual inspections during tubing and connector gluing, verifying glue level, correct assembly of p-cap/luer/sc1/t-cap, tube condition (not stretched, no bruising), absence of contamination, connector integrity, and needle condition (orientation, damage, length).100% functional testing during assembly, including flow tests performed sampling verification under acceptable quality limit (aql) 0.65, including visual and functional tests such as: water leak test static tension tests: tube to connector and tube to luer needle retention flow testing visual inspections for glue curing, absence of bubbles, proper tube location, connector alignment, and protective cap installation.Corrective and preventive action (capa) determination ¿ conclusion: based on the investigation, no further action is required.The record will be closed and monitored through tracking and trending per work instruction (wi) (monthly trips and alerts).Complaint unconfirmed: following investigation, the reported failure could not be confirmed for this complaint.Complaint investigation - conclusion: based on the limited information available, the investigation was completed, and no product malfunction was alleged or identified.The complaint could not be confirmed due to insufficient information, and the record is being closed based on the event description and the information provided in the complaint.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 8021545 manufacturing site: 3003442380.
 
Event or Problem Description
To date, additional patient or event details were received which have been added in h11 (addl mfg narrative).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONTACT DETACH
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
UNOMEDICAL DEVICES S.A DE C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX  88736
Manufacturer (Section G)
UNOMEDICAL DEVICES S.A DE C.V
ave.fomento industrial l9 m3
parque industrial del norte
reynosa,tamps. 88736
MX   88736
Manufacturer Contact
courtney smith
aaholmvej 1 - 3
osted
lejre 4320
DA   4320
6032470255
MDR Report Key25500641
Report Number3003442380-2026-33550
Device Sequence Number18779620
Product Code FPA
UDI-Device Identifier05705244005778
UDI-Public05705244005778
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K160648
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-864
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2026
Supplement Date Received by Manufacturer06/10/2026
Initial Report FDA Received Date06/15/2026
Supplement Report FDA Received Date06/18/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
Patient SexMale
-
-