Based on the available information, this event is deemed to be a reportable serious injury.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Reporting site: 8021545.Manufacturing site: 3003442380.
|
|
Correction: this mdr is being submitted to correct the submitted expiration date under d4 and manufacturing date under h4 and initial report sent to fda under report initial information and initial report sent to fda e4.Additional information - this electronic medical device report (emdr) is being submitted to include the below:
h6: investigation results under type of investigation, investigation findings, investigation conclusions
h11: investigation summary
complaint investigation results
a complaint investigation was conducted based on the event description and the assigned malfunction code no malfunction based on complaint information
based on the available information, the event is consistent with misuse, user-related factors, or no malfunction alleged.In accordance with appendix of work instruction (wi) complaint handling, the need for additional investigation activities, including product testing, or electronic quality management system (eqms) search, was evaluated and determined not to be warranted for this record.However, as a lot number was provided and the event involved a serious injury, a batch record / medical device file review was conducted for lot 6017201 to assess any potential manufacturing-related issues.Batch record / medical device file review:
the batch record / medical device file was reviewed.All required in-process and final tests were completed and met specified requirements.No deviations were identified, and no related maintenance activities were recorded.Conclusion: batch record / medical device file review supports compliance with manufacturing and quality requirements; no issues noted.Complaint unconfirmed:
following investigation the reported failure could not be confirmed for this complaint.Corrective and preventive action (capa) determination assessment - conclusion summary of complaint investigation:
based on the investigation, no further investigation is required.The record will be closed and monitored through tracking and trending per work instruction (wi) (monthly trips and alerts).To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number
reporting site: 8021545
manufacturing site: 3003442380.
|