Model Number 4.23.131
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 02/28/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Implant didn't achieve osseointegration, primary stability was achieved, implant was completely covered with bone, no thread tap used, complication in site preparation, immediate implantation, overheating of bone.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELEMENT RC INICELL, PF 4.5, E 4.2, L6.5
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH  2540
Manufacturer (Section G)
THOMMEN MEDICAL AG
neckarsulmstrasse 28
grenchen 2540
CH   2540
Manufacturer Contact
astrid weber
neckarsulmstrasse 28
grenchen 2540
CH   2540
MDR Report Key25500907
Report Number3003184527-2026-01134
Device Sequence Number21458052
Product Code DZE
UDI-Device Identifier07640156470871
UDI-Public07640156470871
Combination Product (Y/N)N
Initial Reporter CountryES
PMA/510(K) Number
K093615
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number4.23.131
Device Lot Number34339
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/08/2026
Initial Report FDA Received Date06/15/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/13/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age85 YR
Patient SexMale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-