| Catalog Number |
530.710 |
| Medical Device Problem Codes |
Complete Loss of Power (4015); Intermittent Loss of Power (4016)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: check function of device.Check oscillation frequency with frequency meter.During the service evaluation the following failures were identified: functional: intermittent operation.Conclusion: failure reported is not confirmed.Root cause: faulty parts.Action: repair.
|
| |
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the battery oscillator ii for bpl ii device operated intermittently.It was noted in the service order that the device was no working there were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2026.All available information has been disclosed.
|
| |
|
Search Alerts/Recalls
|