| Model Number |
GIF-H180 |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.Date of event is unknown.Additional information will be provided if available.A definitive root cause could not be identified.Based on the results of the investigation, the most probable cause of the complaint was not established.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
The customer returned the gastrointestinal videoscope to the service center, which is an olympus asset with no reported complaint.During the device analysis, the service repair technician indicated white residue in air/water channel and air/water cylinder.There was no patient involvement.
|
| |
|