EIT EMERGING IMPLANT TECHNOLOGIES GMBH EIT TLIF CAGE; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
| Catalog Number |
TUI81428 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Dural Tear (1417)
|
| Date of Event |
05/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint#: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Subject id: (b)(6).Study no: (b)(4).Clinical adverse event received for incidental durotomy repaired during procedure device related: not related, procedure related: probable, date of event: 19 may 2026, date of implant: on (b)(6) 2026, date of revision: no information provided.Device location: l4-l5: yes.Treatment/impact: surgical intervention for the study spine segment (on (b)(6) 2026), depuy synthes products used: catalog number: tui81428, lot number: no information provided, component type: cage, description: eit tlif, h 14mm, 8°, 28/10, catalog number: unk - screws, lot number: no information provided, component type: pedicle screws, description: expedium verse(r) spinal system pedicle screws, catalog number: unk - rods, lot number: no information provided, component type: rods, description: expedium verse(r) spinal system rods.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative:
investigation summary an analysis of the product could not be performed since a physical sample was not received for evaluation.The ri identified no issues during the manufacturing and
release of this device that could have contributed to the problem reported by the customer
as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Based on the information available, there is no indication that a design or manufacturing issue has caused the reported complaint condition, therefore it has been determined that no corrective and preventive action is required at this time.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product is received at a later date, the investigation will be updated as applicable.Device history lot a review of the receiving inspection (ri) for tlif-c ui 14mm 8deg 28/10, was conducted identifying that lot number 455078, was released in one batch.Batch1: lot qty of (b)(4) units were released on sep 9, 2025 with no discrepancies.Supplier : (b)(4).As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device history review the ri identified no issues during the manufacturing and
release of this device that could have contributed to the problem reported by the customer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|