Catalog Number 142290490
Medical Device Problem Code Particulates (1451)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available for evaluation; however, a picture was provided.Investigation is pending.
 
Event or Problem Description
The event involved a secondary set, secure lock, 34 inch.It was reported as small black dots on the internal walls of the drip chambers and has been reported in numerous lots across three product types from icu medical.In some products, particulate matter appears in the lumen or chamber of the tubing.The particulate matter appeared to be embedded in the plastic material but not within the lumen or chamber.No adverse events have been linked to the products as it was noticed prior to patient use and upon inspection of the products, and the biological nature of the particulates remains unknown.The reported model and catalog number was 1422928 and in addition to the black speck, it was also noted that there were some white contaminants present in the product.Sample images provided by the reporter is available in the attachments.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECONDARY SET, SECURE LOCK, 34 INCH
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ICU MEDICAL COSTA RICA LTD.
zona franca global
la aurora heredia
CS 
Manufacturer Contact
reed covert
600 n. field dr.
lake forest, IL 60045
2247062300
MDR Report Key25509013
Report Number9615050-2026-00519
Device Sequence Number10552731
Product Code FPA
UDI-Device Identifier10887787006282
UDI-Public(01)10887787006282(17)280901(10)14524260
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K142367
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number142290490
Device Lot Number14524260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/29/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2026
Supplement Date Received by Manufacturer06/18/2026
Initial Report FDA Received Date06/15/2026
Supplement Report FDA Received Date07/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/20/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-