Catalog Number 2441-0007
Medical Device Problem Code Air/Gas in Device (4062)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/06/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Two sets were returned for investigation.The complaint was unable to be replicated as the sets were able to be primed and all ports were able to be flushed.The root cause could not be determined because the issue could not be replicated.A device history record review was performed.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.
 
Event or Problem Description
It was reported that bd alaris smartsite pump infusion set had air bubbles.The following information was received by the initial reporter with the following: it was reported by customer that a buretrol iv-line set was used, the luer lock port on the buretrol was defective.When syringe was attached to port unable to push in fluid or draw back on syringe.When rn needed to remove air bubbles from the iv tubing unable to put the removed ivig back into the buretro.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD ALARIS SMARTSITE PUMP INFUSION SET
Common Device Name
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS, S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key25511618
Report Number9616066-2026-01566
Device Sequence Number13388330
Product Code FPA
UDI-Device Identifier10885403235047
UDI-Public(01)10885403235047(11)250724(17)280723(10)25075610
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K022209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number2441-0007
Device Lot Number25075610
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/25/2026
Initial Report FDA Received Date06/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/24/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-