• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABIOMED, INC. - 1220648 KIT, 14FR INTRODUCER, 13CM&25CM, STERILE; INTRODUCER, CATHETER (SHEATH)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABIOMED, INC. - 1220648 KIT, 14FR INTRODUCER, 13CM&25CM, STERILE; INTRODUCER, CATHETER (SHEATH) Back to Search Results
Catalog Number 0052-3046
Medical Device Problem Code Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/17/2026
Type of Reportable Event Malfunction
Event or Problem Description
The complainant reported that a patient was implanted with an impella device via the left femoral artery for mechanical circulatory support.It was reported that implantation of an impella cp was aborted due to patient anatomy.No patient harm was reported.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The cause of the difficulty to advance was determined to be patient condition related since the patient had calcium and tortuosity.
 
Additional Manufacturer Narrative
The cause of the difficulty to advance was determined to be patient condition related since the patient had calcium and tortuosity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT, 14FR INTRODUCER, 13CM&25CM, STERILE
Common Device Name
INTRODUCER, CATHETER (SHEATH)
Manufacturer (Section D)
ABIOMED, INC. - 1220648
22 cherry hill drive
danvers MA 01923
Manufacturer Contact
derek sammarco
22 cherry hill drive
danvers, MA 01923
7814934967
MDR Report Key25511811
Report Number1220648-2026-08860
Device Sequence Number10524983
Product Code DYB
UDI-Device Identifier00813502013269
UDI-Public(01)00813502013269(11)251223(10)S10033664
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P140003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0052-3046
Device Lot NumberS10033664
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/12/2026
Supplement Date Received by Manufacturer06/15/2026
Initial Report FDA Received Date06/15/2026
Supplement Report FDA Received Date06/17/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-