| Catalog Number |
0052-3046 |
| Medical Device Problem Code |
Difficult to Advance (2920)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The complainant reported that a patient was implanted with an impella device via the left femoral artery for mechanical circulatory support.It was reported that implantation of an impella cp was aborted due to patient anatomy.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by abiomed inc, or its employees that the report constitutes an admission that the product, abiomed inc, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.The cause of the difficulty to advance was determined to be patient condition related since the patient had calcium and tortuosity.
|
| |
|
Additional Manufacturer Narrative
|
|
The cause of the difficulty to advance was determined to be patient condition related since the patient had calcium and tortuosity.
|
| |
|