C.R. BARD, INC. (BASD) -3006260740 Hickman CV Catheter, Triple-Lumen, 12.5F; Chronic Catheters
|
Back to Search Results |
|
| Catalog Number |
0600650 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Burning Sensation (2146); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/15/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: as the lot number for the device was provided, a review of the device history records will be performed.The sample has not been returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) are adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow up report will be submitted as applicable.Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent a chronic catheter placement procedure using the hickman cv catheter.Post procedure on (b)(6) 2026, the catheter was allegedly found leaked.On (b)(6) 2026, the device was explanted and replaced with another catheter.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2026, a patient underwent chronic catheter placement procedure using the hickman/leonard/broviac central venous catheter.On (b)(6) 2026, during infusion of a vesicant medication through the catheter, serosanguinous drainage consistent with diluted blood was noted leaking from the catheter exit site.Also, patient experienced burning sensation with palpation surrounding the area.The patient experienced a complication that required a subsequent intervention for device removal and replacement.On (b)(6) 2026, the device was explanted.The current status of the patient is unknown.
|
| |
|
Event or Problem Description
|
|
On (b)(6), 2026, a female patient underwent chronic catheter placement procedure using the hickman/leonard/broviac central venous catheter.On (b)(6), 2026, during infusion of a vesicant medication through the catheter, serosanguinous drainage consistent with diluted blood was noted leaking from the catheter exit site.Also, the patient experienced burning sensation with palpation surrounding the area.The patient experienced a complication that required a subsequent intervention for device removal and replacement.On (b)(6), 2026, the device was explanted.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: based on the additional information received from the canada quality, indicating that while a vesicant medication was infusing through the catheter, serosanguinous drainage (diluted blood) was leaking from the exit site and the patient experienced a burning sensation upon palpation of the surrounding area, all infusions were immediately stopped.Upon reassessment of the event and considering this additional clinical information, the case was upgraded to a serious injury, and the file was reassessed for reportability and determined to be reportable as serious injury.Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: one 12.5fr hickman t/l catheter was returned for evaluation.Visual, microscopic, tactile and functional evaluations were performed on the returned device.An in-house syringe was connected to the blue, red, and white catheter hubs.Each lumen was patent for both infusion and aspiration, with no leaks observed during testing.Hydrostatic pressure testing was performed by clamping the distal end of the catheter while infusing fluid.An in-house syringe filled with dyed water was attached to each catheter hub, and no leakage was observed from any hub during infusion.Therefore, the investigation is inconclusive for the reported fluid leak issue as the sample evaluation results indicating no leak and difficulty/issue under laboratory conditions may not be representative of the condition of the device during use.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.B1, b2, b5, h1, h6 (patient, method, result, conclusion).Section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this lot.Investigation summary: one 12.5 (b)(6) catheter was returned for evaluation.Visual, microscopic, tactile and functional evaluations were performed on the returned device.An in-house syringe was connected to the blue, red, and white catheter hubs.Each lumen was patent for both infusion and aspiration, with no leaks observed during testing.Hydrostatic pressure testing was performed by clamping the distal end of the catheter while infusing fluid.An in-house syringe filled with dyed water was attached to each catheter hub, and no leakage was observed from any hub during infusion.Therefore, the investigation is inconclusive for the reported fluid leak issue as the sample evaluation results indicating no leak under laboratory conditions may not be representative of the condition of the device during use.A definitive root cause could not be determined based upon the provided information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.G3, h.4.Device manufacture date: unknown
section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|