This event was initially deemed not reportable based on our reporting strategy active at the time of the alert date; however, following a recent retrospective review of complaints this event was made reportable out of an abundance of caution.A review of the device history record is not possible as no lot number was provided; therefore, the udi-pi is unavailable.The actual complaint product was returned for evaluation.The tube was also flushed through the y connector (proximal end).No resistance was felt while flushing and the liquid was seen exiting out where the burst in the tube was located which is below the y-port connector.At the burst location, the tubing appeared to have kinked then burst to create an opening.Both breaking points did not exhibit dried foreign material residue.However, when feeling the tubing further down, at 60cm marking, hardened feeling was felt on this area.The tubing was cut and opened to inspect the inner surface of tubing walls.When opened and inspected, orange/ red substance residues were stuck in the tube.A root cause has not been established.All information reasonably known as of 16 jun 2026 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the reported event but is relaying the information provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
|
Avanos medical inc.Received a single report that referenced three different incidents, which were associated with separate units, involving three different patients.This is the first of three reports.Refer to for the first report, refer to 9611594-2026-00450 for the second report , refer to 9611594-2026-00451 for the third report.It was reported that a nasogastric feeding tube was placed.It clogged and was removed on (b)(6) 2024.Kclor cr given.
|