• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. COATED VICRYL; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. COATED VICRYL; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP695H
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/29/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Additional information provided: no patient involvement patient status/ outcome / consequences- no, was other medical intervention (e.G.X-rays, additional procedures, prescriptions, otc, revision) required: no, is the patient part of a clinical study-unknown, device property of-none, device in possession of- none, additional information has been requested however not received.Attempts to obtain the device have been made.If further details are received at a later date a supplemental medwatch will be sent.How was the product purchased? is there an indication of how the product was distributed? is there any indication of the source? based on the packaging, is there any indication of which market the original genuine product may have come from? please provide the source or name and title of external person providing answers to follow-up (external person submitting answers to sales rep or loc/bq).A manufacturing record evaluation was performed for the finished device lot number, and no related non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
No complaint was logged by hospital - i emailed our quality department with an observation and she asked me to log a complaint.The color of the suture is not displayed on the packaging of the following two sutures.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COATED VICRYL
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key25520819
Report Number2210968-2026-06385
Device Sequence Number21484396
Product Code GAM
UDI-Device Identifier10705031054813
UDI-Public(01)10705031054813(17)280630(10)10A2CJ
Combination Product (Y/N)Y
Initial Reporter CountrySF
PMA/510(K) Number
K181652
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberVCP695H
Device Lot Number10A2CJ
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/29/2026
Supplement Date Received by Manufacturer06/18/2026
Initial Report FDA Received Date06/16/2026
Supplement Report FDA Received Date07/06/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/31/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-