• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD SLIMLINE SIS EZ; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD SLIMLINE SIS EZ; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 0644-021-01
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/02/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a procedure, the fiber got burnt.The procedure was completed with another fiber without patient complications.
 
Additional Manufacturer Narrative
Upon receipt of this fiber at our quality assurance laboratory, this device was thoroughly analyzed.Visual analysis of the returned fiber found that the fiber was received with no defects.Microscope inspection of the fiber determined that the fiber tip was degraded with burn marks on fiber jacket; also, the connector was burnt and there was no light exiting the connector face of the fiber, indicating a connector fracture.Functional test found that the aiming beam was weak.The connector fractured could lead to an insufficient aiming beam or low laser energy output and up to a complete inability for the fiber to fire.A fracture within the connector could occur due to procedural handling of the fiber during setup or use.Also, the interaction between the console and the fracture connector could lead to an overheating of the fiber, that could cause the fiber separated from the connector.Therefore, the reported event of fiber overheated was confirmed.The device history record confirmed that the device met with all manufacturing specifications prior to distribution and there were no manufacturing deviations.The instructions for use (ifu) were reviewed and did not reveal any evidence of device off-label use or failure to follow instructions.The ifu states: inspect the fiber for kinks, punctures, fractures, or other damage.If the fiber appears damaged, do not use the fiber; return it to the supplier for replacement and hold the fiber tip to a non-reflective surface, and ensure that a circular red/green spot appears.If the spot is not circular, cleave the fiber.If the spot is weak or not visible, discard the fiber or return it to the supplier for replacement.A risk review was completed and confirmed that the event of fiber overheating of device or device component is defined in the risk documentation.Based on the analysis performed, a conclusion code of cause traced to component failure was assigned to this investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLIMLINE SIS EZ
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidma 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
MOLEX LLC
18019 n 25th ave
phoenix AZ 85023
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key25520917
Report Number2124215-2026-31915
Device Sequence Number13265404
Product Code GEX
UDI-Device Identifier07290109145600
UDI-Public(01)07290109145600(17)281106(10)15141223
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0644-021-01
Device Catalogue Number0644-021-01
Device Lot Number0015141223
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/24/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2026
Initial Report FDA Received Date06/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/08/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexMale
Patient Weight80 KG
-
-