| Catalog Number |
786624 |
| Medical Device Problem Code |
Calcified (1077)
|
| Health Effect - Clinical Codes |
Exposure to Body Fluids (1745); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/09/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the patient had the ureteral stent in place for less than two months, and severe stone encrustation was found, so the stent was removed and a new stent was implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.(b)(6).Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient had the ureteral stent in place for less than two months, and severe stone encrustation was found, so the stent was removed and a new stent was implanted.The user was exposed to hazardous substances, blood, or body fluids.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation determined the product does not meet specifications, and no new product nonconformance was identified.Visual inspection of the one (1) received photo sample was performed upon intake.Photo #1 depicts calcification present on the stent.The labeling is found to be adequate.Dhr review for corporate lot no.Ngjt1273 catalog number 786624 (bard inlay optima ureteral stent, hydrophobic coated, open end with suture loop 6fr x 24cm with nitinol guidewire) used in the manufacture of the reported product did not show any rejects, planned deviations, reworks or component quality issues.As per investigation results, the annex g code was found to be g01-biological and chemical.Corrections made to tab(s) d, f, h
initial reporter name: the second hospital of zhejiang university, school of medicine
udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|