• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SINGLE-USE DIRECTVIEW MAC S4; LARYNGOSCOPE, RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VERATHON MEDICAL ULC GLIDESCOPE SPECTRUM SINGLE-USE DIRECTVIEW MAC S4; LARYNGOSCOPE, RIGID Back to Search Results
Model Number 0574-0188
Medical Device Problem Codes Erratic or Intermittent Display (1182); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A customer reported the screen on their glidescope avl system went black when the glidescope spectrum single-use directview mac s4 laryngoscope was inserted into the patient's mouth and returned to normal when removed.The facility's biomedical engineering department reported being unable to replicate the issue using a different glidescope blade.No delay in procedure, use of a backup device, or harm to the patient was reported.The incident date is unknown.
 
Additional Manufacturer Narrative
The customer's glidescope spectrum single-use directview mac s4 was not returned to verathon for evaluation, therefore the cause could not be determined.Verathon has made multiple follow-up attempts to the customer for additional event information.To date, no response has been received.Review of complaint history for the device's lot number was not able to be performed as no lot number was provided.Trending analysis for glidescope spectrum single-use laryngoscopes does not identify any trends exceeding acceptable limits.Verathon confirmed that the risk associated with the hazardous scenario has been adequately captured and characterized per the system risk assessment.Corrective action is not required at this time.Verathon will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE SPECTRUM SINGLE-USE DIRECTVIEW MAC S4
Common Device Name
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key25522209
Report Number9615393-2026-00072
Device Sequence Number13400153
Product Code CCW
UDI-Device Identifier00879123006257
UDI-Public(01)00879123006257
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0574-0188
Device Catalogue Number0270-0933
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/19/2026
Initial Report FDA Received Date06/16/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-