| Catalog Number |
SSF-650 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Patient reported ¿encapsulated¿.Later healthcare professional reported ¿rupture¿.Later patient reported "nipple discharge".This record is for the right side.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Patient reported ¿encapsulated¿.Later healthcare professional reported ¿rupture¿.Later patient reported ""nipple discharge".This record is for the right side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: d9, h3, h6.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
rupture: not observed through visual inspection.Nipple discharge: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis deformation, creases and wear abrasion are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|