| Model Number |
VTICM5 13.2 (-14.00/2.00) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
04/30/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturer narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim#: (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced excessive vaulting with significant reduction of irido-corneal angles, refractive surprise, angle closure, blurry and cloudy vision.The lens was exchanged with a shorter length lens.The problem was resolved.The cause of the event was reported as other and noted "issue with measurements for diameter".
|
| |
|