• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION OverStitch 2-0 Polypropylene Suture; Endoscopic Tissue Approximation Device

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION OverStitch 2-0 Polypropylene Suture; Endoscopic Tissue Approximation Device Back to Search Results
Model Number M00505451
Medical Device Problem Codes Break (1069); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 05/21/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported to boston scientific corporation that an overstitch suture was used in the stomach during an endoscopic sleeve gastroplasty (esg) procedure on (b)(6) 2026.During the procedure, the physician was placing a u-stitch suture pattern when two sutures broke in the middle of the suture material rather than detaching from the anchor.The first affected suture was removed from use and replaced with a new device; however, the issue recurred with a second suture, requiring another replacement device to be opened.Approximately 10 minutes were required each time to cut and remove the broken suture, resulting in an estimated 20-minute procedure prolongation.The procedure was ultimately completed using another of the same device from a different lot.During the procedure, the patient developed a hematoma located in the body of the stomach between the greater and lesser curvature.No additional intervention was performed to treat the hematoma; the physician sutured around the affected area and completed the procedure.The physician was unable to determine whether the device or the procedure caused or contributed to the hematoma.The patient was not admitted beyond the standard of care, subsequently recovered, and was discharged home.This is the second of two overstitch sutures used in the same procedure.
 
Additional Manufacturer Narrative
Block h6: device code a0401 is being used to capture the reportable event of suture break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OverStitch 2-0 Polypropylene Suture
Common Device Name
Endoscopic Tissue Approximation Device
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
Boston Scientific Corporation
2546 calle primera
alajuela
CS  
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
+1(914)571-6918
MDR Report Key25524911
Report Number3005099803-2026-02494
Device Sequence Number21519243
Product Code OCW
UDI-Device Identifier00191506060138
UDI-Public(01)00191506060138(17)301001(10)37603787
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/14/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00505451
Device Catalogue NumberM00505451
Device Lot Number0037603787
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer06/11/2026
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2026
Supplement Date Received by Manufacturer07/17/2026
Initial Report FDA Received Date06/16/2026
Supplement Report FDA Received Date08/14/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/02/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-