Model Number M00558380
Medical Device Problem Code Burst Container or Vessel (1074)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/01/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf device code a0402 captures the reportable event of balloon burst.Block h11: investigation results the device was not returned for analysis and was reported to be disposed; therefore, a technical analysis could not be performed.With all available information, boston scientific concludes the reported event of balloon burst could not be confirmed.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause not established.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of balloon burst was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was attempted to be used in the esophagus during an esophagogastroduodenoscopy (egd) procedure for dysphasia performed on an unknown date.During the procedure, the balloon burst after reaching a diameter of 20mm.The procedure was completed with another of the same device.No piece of balloon detached inside the patient.There were no patient complications reported as a result of this event.The patients condition at the conclusion of the procedure was reported to be fully recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE FIXED WIRE
Common Device Name
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key25524941
Report Number3005099803-2026-02491
Device Sequence Number10741412
Product Code KNQ
UDI-Device Identifier08714729202011
UDI-Public(01)08714729202011(17)280226(10)38702266
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 06/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM00558380
Device Catalogue NumberM00558380
Device Lot Number0038702266
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/21/2026
Initial Report FDA Received Date06/16/2026
Date Device Manufactured02/26/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-