| Model Number |
22550R |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
03/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that after cardioverting a patient, the staff/physician noticed a visible spark coming from the pads, smell of smoke, and patient had a burn on their skin in the shape of one of the black inner-ring corners.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|