| Model Number |
500001002001 |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of the complaint history for sn (b)(6) was performed in salesforce which did not locate similar complaint(s) with the same failure mode for this serial number.A review of the device history record for sn (b)(6) was performed from the date of manufacture, 30-dec-2019 and confirmed that this device was not previously returned for servicing and there were no production failures which correlates to the customer reported issue.Upon investigation of the actual device used in this incident, it was determined that the battery was flashing out and beeping.A field service engineer removed the backup uninterruptible power supply (ups) and connected the console directly to an emergency wall outlet, then replaced the uninterruptible power supply (ups) package.The system functioned as intended after the field service engineer repaired the device.
|
| |
|
Event or Problem Description
|
|
It was reported that when using the bd pyxis¿ medstation¿ 4000 console battery started flashing, produced a beeping sound and went offline on remote support service.The customer stated that this malfunction occurred while dispensing the medication and the user managed to get the medication from other sources, which caused a delay to the patient care.There were no adverse events or injuries reported based on this event.
|
| |
|