| Catalog Number |
15-397 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Lab analysis: analysis of the returned device identified: weight to spec and deformation device.No other observation observed.Base on the device analysis the final assessment is: no issues found related with the manufacturing process.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture grade unknown".This record is for the left side.Device was explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported "capsular contracture grade unknown".This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
This report was impacted by emdr scheduled maintenance occurring july 22-27, 2026.Device evaluation:
per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required:
¿ capsular contracture: unable to observe through visual inspection as it is a physiological phenomenon.None of the other observations performed during the device analysis (deformation) are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.Additional, changed, and/or corrected data: h11.
|
| |
|