• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO VILLALBA EPSILA EV¿ MRI SURESCAN¿; PERMANENT DEFIBRILLATOR ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO VILLALBA EPSILA EV¿ MRI SURESCAN¿; PERMANENT DEFIBRILLATOR ELECTRODES Back to Search Results
Model Number EV240152
Medical Device Problem Code Signal Artifact/Noise (1036)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/22/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during the implant procedure, when the extravascular (ev) lead was connected to the analyzer cables, r-waves we re appropriately measured.After the physician pulled the lead back slightly and after reconnecting there was nothing but noise on the analyzer.The cables were reconnected and the same result occurred.A second pair of analyzer cables was tried with the same result.The physician took a second lead and implanted it with no issues noted.The initial ev lead was not implanted.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the full lead was returned, analyzed, and no anomalies were found.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EPSILA EV¿ MRI SURESCAN¿
Common Device Name
PERMANENT DEFIBRILLATOR ELECTRODES
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key25536682
Report Number2649622-2026-18041
Device Sequence Number18577977
Product Code NVY
UDI-Device Identifier00763000369767
UDI-Public(01)00763000369767(17)280202
Combination Product (Y/N)N
Initial Reporter StateRI
Initial Reporter CountryUS
PMA/510(K) Number
P220012
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberEV240152
Device Catalogue NumberEV240152
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/22/2026
Supplement Date Received by Manufacturer06/18/2026
Initial Report FDA Received Date06/17/2026
Supplement Report FDA Received Date07/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/04/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age45 YR
Patient SexMale
Patient Weight84 KG
-
-