(b)(4).Date sent: 6/17/2026.D4: batch #596d86.Investigation summary: the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that one psee60a device was returned with no reload present.During visual inspection, the bailout door was noted to be out of position; additionally, the bailout lever was damaged, likely due to interaction with the cam bailout.It should be noted after the manual override system is used the instrument is disabled and cannot be used for any subsequent firings.The bailout system was reset and then, it was tested for functionality in the straight position with a test reload and achieved its complete firing sequence without any difficulties.The staple line and cut line were complete and the staples met the staple release criteria.It is possible that an outside the normal use force was applied on the lever bailout as it was pushed down once the device was bailout.The event described of difficult to open could not be confirmed since the device opened as expected.As part of ethicon endo surgery¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device psee45a lot number a9861x and no non-conformances were identified.A manufacturing record evaluation was performed for the finished device batch number 596d86, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|