| Model Number |
P7900A |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The baxter technician found the auxiliary power cord needed to be replaced.A search of the baxter maintenance records showed baxter performed preventative maintenance on this bed in mar 13, 2025.It is unknown if the facility performed any other preventative maintenance on this bed.Per the baxter service manual, it is necessary for the centrella bed to have an effective maintenance program.We recommend that you do annual preventive maintenance.Make sure the power cord is not damaged and it is connected to the bed.Replace the power cord as necessary.Check the ac inlet to psm cable assembly for kinks, damage, and connections.Replace or connect the cable as necessary.Replace parts as necessary.The technician replaced the auxiliary power cord to resolve the reported event.Based on this information, no further action is required.
|
| |
|
Event or Problem Description
|
|
Baxter received a report that centrella frame had auxiliary power cord damaged with copper wires exposed.This occurred before use.There was no patient/user injury, medical intervention, symptom, or adverse event reported.
|
| |
|