| Model Number |
6640440 |
| Medical Device Problem Code |
Failure of Device to Self-Test (2937)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
It was reported that the ventilator failed the internal leakage test during the pre-use check.There was no patient involvement.A third-party field service engineer (fse) has investigated the reported ventilator on-site.The fse replaced the air gas module to resolve the issue.Following the replacement, the ventilator passed all functional and safety tests and was cleared for clinical use.The device's logs were not provided for further analysis.The air gas module regulates the inspiratory gas flow and a faulty air gas module may affect the delivered volume or gas mixture (o2/air).The cause of the reported issue in this customer product complaint has been determined.The issue was related to the air gas module.The defective part was not returned for investigation despite our request.According to the information received, it was retained by the customer in accordance with their policy.
|
| |
|
Event or Problem Description
|
|
It was reported that the ventilator failed the internal leakage test during the pre-use check.There was no patient involvement.Manufacturer's ref.#: (b)(4).
|
| |
|