| Model Number |
305U225 |
| Medical Device Problem Codes |
Calcified (1077); Gradient Increase (1270); Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Codes |
Calcium Deposits/Calcification (1758); Aortic Valve Insufficiency/ Regurgitation (4450)
|
| Date of Event |
05/27/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that 15 years post implant of this 25mm aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported as increased gradient with severe aortic regurgitation.Evidence of calcification was seen in the left ventricular outflow tract (lvot) under left coronary cusp (lcc).Subsequently on (b)(6) 2026 the patient successfully had aortic transcatheter valve-in-valve replacement.No additional adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
Additional information received reported the value of gradient was 28mmhg and peak velocity of 3.42m/s.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B5 updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|